The DE-916 High-performance Isolated Dynamic Signal Test and Analysis System is specially designed for accurate capture of transient signals in a strong interference environment. By means of multi-layer shielding technique, DE-916 effectively eliminate electromagnetic radiation and conduction interference, and ensure the accurate measurement of force, pressure, stress, strain, displacement, velocity, acceleration and other physical quantities.
Modular design, expandable to over 500 channels via Ethernet
Independent 16-bit SAR A/D per channel, multiple channel parallel operation, no interaction between channels
Built-in bridge completion and Excitation
Multi-layer shielding and advanced isolation technology, electromagnetic noise and conducted interference eliminated effectively
Up to 20MH/channel for transient sampling rate and 1MHz/channel for continuous sampling rate
CMRR over 120dB
Built-in 24V/4mA biasing circuit, used to acquire the output signal of IEPE acceleration sensors and microphones
Built-in Butterworth low-pass filter
DMA real-time data transfer for ensuring the high-speed and stable transmission of data without drop-out
Suport EID and TEDS
Input protection and signal state indication
Number of Input Channel | 1 channle/card, 8 card slots (19" 3U chassis) |
Input Coupling | GND, Dif-DC, Sin-DC, AC, IEPE |
Input Impedance | 10MΩ/10pF |
Full-Scale Voltage Value | ±0.01V, ±0.02V, ±0.05V, ±0.1V, ±0.2V, ±0.5V, ±1.0V, ±5.0V, ±10.0V |
Indication Error | <0.1% of F.S. |
Stability | <0.05%/24h |
Nonlinearity | ≤0.01% of F.S. |
Distortion | ≤0.5% |
Noise | ≤8μVRMS |
Zero Drift | <3μV/8h |
CMV | ±500V (DC/AC peak value) |
CMR | ≥120dB |
Full-scale Strain Value | ±1000με, ±10000με, ±100000με |
Bridge Excitation | |
Bridge Configuration | Full, half, three-wire quarter bridge |
Bridge Completion Resistors | 120Ω/350Ω (Three-wire quarter bridge) 50Ω ~ 10000Ω (Half bridge / Full bridge) |
Bridge Voltage | 2V, 5V, 10V, 24V DC within 0.01% |
Current | Max. 50mA |
LPF | |
Transfer Characteristic | Butterworth low-pass filter |
Cut-off requency (-3dB±1dB) | 300Hz, 1kHz, 3kHz, 10kHz, 30kHz, 100kHz, 300kHz, PASS |
Flatness | <0.1dB Within 2/3 of cutoff frequency |
Stop-band Attenuation | - 18dB/Oct. |
Communication | Gigabit Ethernet |
A/D Converter | 16-bit SAR |
Freq. Response | DC ~ 1MHz (+0.5dB ~ -3dB) |
Max. Continuous Sampling Rate | 1MHz |
Max. Transient Sampling Rate | 20MHz |
Power Supply | 100 ~ 240V AC, 50-60Hz, 100W |
Dimensions | 482×133×338 mm |
Weight | Approx. 20.5kg |
Environmental Conditions | |
Operating Temperature | - 10°C to 50°C |
Operating Humidity | 20 ~ 90%RH@40°C |
Storage Temperature | - 40°C to 60°C |
Storage Humidity | 90%RH24h@50°C |
Vibration | Frequency cycle range: 5Hz ~ 55Hz ~ 5Hz Drive amplitude (peak): 0.19mm Sweep frequency: ≤1Oct./minDuration of resonant: 10min Vibration direction: x, y, z |
Figure 1 Single system block diagram
Figure 2 Multi-system Block Diagram
DE-BPS Basic Platform Software
Running on XP/Win7/Win8/Win10 operating system
Parameters setting, Function control, Real-time/post-acquisition analysis, data browsing, cursor readouts, scaling curve, data management and simple processing, report generation, long-term continuous data recording, etc.
DE-916 DAQ Card | |
| |
DE-916 Host Chassis | |
| |
DT5857-8 Charge Conditioner (Optional) | |
| |
DT3811-8 Current Loop Conditioner (Optional) | |
| |
DT3814-8 Thermistor Conditioner (Optional) | |
| |
DE-9-S Signal Source Module (Optional) | |
| |
TM03 Tachometer (Optional) | |
| |
DT5611A Synchronization Box (Optional) | |
| |
C-1 Cable | |
|